PRESS-NEWS.org - Press Release Distribution
PRESS RELEASES DISTRIBUTION

Cardiac device wearers should keep distance from smartphones

Patients should hold phone to opposite ear and not store phone in pocket over device

2015-06-22
(Press-News.org) Milan, Italy, June 22 -- Cardiac device wearers should keep a safe distance from smartphones to avoid unwanted painful shocks or pauses in function, reveals research presented today at EHRA EUROPACE -- CARDIOSTIM 2015 by Dr. Carsten Lennerz, first author and cardiology resident in the Clinic for Heart and Circulatory Diseases, German Heart Centre, Munich, Germany.

The joint meeting of the European Heart Rhythm Association (EHRA) of the European Society of Cardiology (ESC) and Cardiostim is being held in Milan, Italy. The scientific programme is here: http://www.flipsnack.com/Escardio/ehra-europace-cardiostim-2015-advance-programme.html.

Lennerz said: 'Pacemakers can mistakenly detect electromagnetic interference (EMI) from smartphones as a cardiac signal, causing them to briefly stop working. This leads to a pause in the cardiac rhythm of the pacing dependent patient and may result in syncope. For implantable cardioverter defibrillators (ICDs) the external signal mimics a life threatening ventricular tachyarrhythmia, leading the ICD to deliver a painful shock.'

Device manufacturers and regulatory institutions including the US Food and Drug Administration (FDA) recommend a safety distance of 15 to 20 cm between pacemakers or ICDs and mobile phones. The advice is based on studies performed primarily in pacemakers 10 years ago. Since then smartphones have been introduced and mobile network standards have changed from GSM to UMTS and LTE. New cardiac devices are now in use including ICDs, cardiac resynchronisation therapy (CRT) and MRI compatible devices.

The current study evaluated whether the recommended safety distance was still relevant with the new smartphones, networks and cardiac devices. A total of 308 patients (147 pacemakers and 161 ICDs, including 65 CRTs) were exposed to the electromagnetic field of three common smartphones (Samsung Galaxy 3, Nokia Lumia, HTC One XL) which were placed on the skin directly above the cardiac device.

The smartphones were connected to a radio communication tester, which works like a mobile network station. The investigators put the smartphones through a standardised protocol of the calling process which included connecting, ringing, talking and disconnecting. The actions were performed in GSM, LTE and UMTS at the maximum transmission power and at 50 Hz, a frequency known to influence cardiac implantable electronic devices. Electrocardiograms (ECGs) were recorded continuously and checked for interference.

Lennerz said: 'From earlier studies we know that the most vulnerable phases of a call are ringing and connecting to the network, not talking, so it was important to analyse these separately.' More than 3,400 tests on EMI were performed. One out of 308 patients (0.3 percent) was affected by EMI caused by smartphones. This patient's MRI compatible ICD misdetected electromagnetic waves from the Nokia and HTC smartphones operating on GSM or UMTS as intracardiac signals.

Lennerz said: 'Interference between smartphones and cardiac devices is uncommon but can occur so the current recommendations on keeping a safe distance should be upheld. Interestingly, the device influenced by EMI in our study was MRI compatible which shows that these devices are also susceptible.'

Professor Christof Kolb, last author and head of the Department of Electrophysiology at the German Heart Centre, said: 'Nearly everyone uses smartphones and there is the possibility of interference with a cardiac device if you come too close. Patients with a cardiac device can use a smartphone but they should not place it directly over the cardiac device. That means not storing it in a pocket above the cardiac device. They should also hold their smartphone to the ear opposite to the side of the device implant.'

In a second study on EMI, researchers advise limiting exposure to high voltage power lines.2 The study was conducted in response to public concerns about bicycle routes and walking paths under high voltage power lines (230 kV and more) and whether these are safe for patients with cardiac devices. These high electric fields are also encountered in utility substations where employees who bring up power lines, conduct maintenance or work within the buildings (cleaners, for example) may be exposed.

Dr. Katia Dyrda, a cardiologist at Montreal Heart Institute, University of Montreal, said: 'High electric fields may interfere with the normal functioning of cardiac devices, leading to the withholding of appropriate therapy (anti-bradycardia pacing, for example) or to the delivery of inappropriate shocks. The International Organization for Standardization says pacemakers and ICDs should give resistance up to 5.4 kV/m (for 60 Hz electric fields) but electric fields can reach 8.5 kV/m under high voltage power lines and 15 kV/m in utility substations.'

She added: 'There is a lot of interest in using the areas under power lines as bicycle paths or hiking trails because it's essentially free space. But patients and the medical community want to understand the risks. There are no recommendations from device manufacturers about power lines or higher electric fields.' The study exposed 40 cardiac devices (21 pacemakers and 19 ICDs) from five manufacturers to electric fields up to 20 kV/m in a high voltage laboratory. The devices were mounted in a saline tank at human torso height. Devices were set up as both left and right sided pectoral implants.

The researchers found that when pacemakers were programmed to nominal parameters and in bipolar mode they were immune to EMI up to 8.6 kV/m. But when programmed to higher sensitivity levels or in unipolar mode, the EMI threshold decreased to as low as 1.5 kV/m in some devices. When programmed to nominal parameters, all ICDs were immune to EMI up to 2.9 kV/m . There was no difference in EMI thresholds between left and right sided implants.

Dyrda said: 'There is no significant concern for patients with pacemakers programmed in the usual configuration (nominal settings, in bipolar mode). For the minority of patients with devices in unipolar mode or with very sensitive settings, counselling should be given at implantation or at medical follow-up.'

She added: 'There is no need for patients with a pacemaker or ICD to avoid crossing under high voltage power lines (> 230 kV) but patients should avoid staying in a stationary position underneath them. Passing near pylons rather than between two pylons mitigates exposure to the electric field because the wires sag in the middle and the field is higher at this location.'

Dyrda emphasised that this advice does not concern distribution lines (lines delivering electricity to homes), as the 60 Hz electric field that they generate is very low. She added: 'Patients ask us if they should avoid driving on roads that cross under high voltage power lines. The answer is no. If you're in a vehicle you are always protected because your car acts as a Faraday cage and shields you automatically.' Employees with a pacemaker or defibrillator should tell their employer so that their safety at work can be carefully evaluated, urged Dyrda. She said: 'Our study tested the effect of electric fields up to 20 kV/m and the results can be used to assess individual risks depending on exposure levels during specific tasks and the type and model of cardiac device. This may lead to job adjustments or, more rarely, to a job change.'

INFORMATION:

Notes to editor

1Dr. Carsten Lennerz will present the abstract 'Electromagnetic interference between smartphones and current cardiac implantable electronic devices: rare but present' during Poster session 3: Devices (Pacing) on 22 June at 08:30 in the Poster Area: http://spo.escardio.org/SessionDetails.aspx?eevtid=1087&sessId=16282&subSessId=4329&searchQuery=%2fdefault.aspx%3feevtid%3d1087%26days%3d%26topics%3d%26types%3d%26rooms%3d%26freetext%3dlennerz%26sort%3d1%26page%3d1%26showResults%3dTrue%26nbPerPage%3d20%26WithWebcast%3d%26WithSlides%3d%26WithAbstract%3d%26WithReport%3d%26scroll%3D0#.VWQd3E9Viko

2Dr. Katia Dyrda will present the abstract 'Interference resistance of pacemakers and defibrillators to 60 Hz electric fields' during Poster session 5: Devices (Pacing) on 23 June at 08:30 in the Poster Area: http://spo.escardio.org/SessionDetails.aspx?eevtid=1087&sessId=16284&subSessId=4393&searchQuery=%2fdefault.aspx%3feevtid%3d1087%26days%3d%26topics%3d%26types%3d%26rooms%3d%26freetext%3ddyrda%26sort%3d1%26page%3d1%26showResults%3dTrue%26nbPerPage%3d20%26WithWebcast%3d%26WithSlides%3d%26WithAbstract%3d%26WithReport%3d%26scroll%3D0#.VWQe6U9Viko

About EHRA EUROPACE -- CARDIOSTIM

EHRA EUROPACE -- CARDIOSTIM is an established, international conference attracting key opinion leaders, well-recognised scientists, physicians, allied professionals and industry.

About the European Heart Rhythm Association (EHRA)

The European Society of Cardiology (ESC) represents more than 80 000 cardiology professionals across Europe and the Mediterranean. Its mission is to reduce the burden of cardiovascular disease in Europe. The European Heart Rhythm Association (EHRA) is a registered branch of the ESC. Its aim is to improve the quality of life of the European population by reducing the impact of cardiac arrhythmias and reduce sudden cardiac death.

Information for journalists attending EHRA EUROPACE -- CARDIOSTIM 2015

EHRA EUROPACE -- CARDIOSTIM 2015 will be held 21 to 24 June in Milan, Italy at the MiCo Milano Congress. The full scientific programme is available here: http://www.flipsnack.com/Escardio/ehra-europace-cardiostim-2015-advance-programme.html To register onsite please bring a valid press card or appropriate letter of assignment with proof of three recent published articles (cardiology or health-related, or referring to a previous ESC Event).

Press registration is not available to Industry or its Public Relations representatives, event management, marketing or communications representatives.



ELSE PRESS RELEASES FROM THIS DATE:

Patients test drive pacemaker before choosing permanent implant

2015-06-22
Milan, Italy - 22 June 2015: Patients are test driving a pacemaker outside the skin before deciding whether to have a permanent implant, reveals novel research presented today at EHRA EUROPACE - CARDIOSTIM 2015 by Professor Michael Giudici, director of arrhythmia services in the Department of Internal Medicine, University of Iowa Hospitals and Clinics, Iowa City, US.1 "Wow, I didn't realise how terrible I felt. This is great. When can we schedule my pacemaker?" The joint meeting of the European Heart Rhythm Association (EHRA) of the European Society of Cardiology (ESC) ...

Study examines US trends in the use of penile prostheses to treat erectile dysfunction

2015-06-22
US researchers have published the first large population-based study of nationwide trends in erectile dysfunction and its surgical management from 2001 to 2010. Using data from Medicare beneficiaries, the investigators found that while the prevalence of erectile dysfunction increased by 165%, the use of penile prostheses decreased by 50%, from 4.6% in 2001 to 2.3% in 2010. Prostheses have been increasingly used in sicker patients with significant comorbidities, however. Age, ethnicity, and geography often influenced whether patients would undergo penile prosthetic placement. "The ...

Study finds decreased rates of high-grade cervical lesions in young women

2015-06-22
A new analysis indicates that rates of high-grade cervical lesions decreased in young U.S. women after vaccines were made available to protect against human papillomavirus (HPV), but the trend may be due in part to changes in cervical cancer screening recommendations. Published early online in Cancer, a peer-reviewed journal of the American Cancer Society, the study illustrates some of the challenges in monitoring a vaccine's impact during a time of concurrent changes in screening. More than half of high-grade cervical lesions, which are abnormal tissue changes that can ...

How to predict biphasic allergic reactions in children

2015-06-22
Ottawa, Ontario, June 22 -- Children are more likely to have a repeat, delayed anaphylactic reaction from the same allergic cause, depending on the severity of the initial reaction. The first pediatric study to look at the predictors for this phenomenon was published today in Annals of Allergy, Asthma & Immunology. Anaphylaxis is a severe, allergic reaction that is rapid in onset and can result in death. Some children are at risk of delayed ('biphasic') anaphylactic reactions. Delayed reactions occur when the initial symptoms of allergic reaction go away but then return ...

Silent flights: How owls could help make wind turbines and planes quieter

2015-06-22
An investigation into how owls fly and hunt in silence has enabled researchers to develop a prototype coating for wind turbine blades that could significantly reduce the amount of noise they make. Early tests of the material, which mimics the intricate structure of an owl's wing, have demonstrated that it could significantly reduce the amount of noise produced by wind turbines and other types of fan blades, such as those in computers or planes. Since wind turbines are heavily braked in order to minimise noise, the addition of this new surface would mean that they could ...

Children with good memories are better liars, research shows

2015-06-20
Children who benefit from a good memory are much better at covering up lies, researchers from the University of Sheffield have discovered. Experts found a link between verbal memory and covering up lies following a study which investigated the role of working memory in verbal deception amongst children. The study saw six to seven year old children presented with the opportunity to do something they were instructed not to -peek at the final answers on the back of a card during a trivia game. A hidden camera and correct answers to the question, which was based on ...

Moffitt researchers discover mechanism leading to BRAF inhibitor resistance in melanoma

2015-06-19
TAMPA, Fla. - The development of targeted therapies has significantly improved the survival of melanoma patients over the last decade; however, patients often relapse because many therapies do not kill all of the tumor cells, and the remaining cells adapt to treatment and become resistant. Moffitt Cancer Center researchers have discovered a novel mechanism that can lead melanoma cells to develop resistance to drugs that target the protein BRAF. Mutations in the gene BRAF are the most common mutation found in melanoma, with up to 50 percent of tumors testing positive for ...

Internists give Senate recommendations to improve care for patients with chronic diseases

2015-06-19
The American College of Physicians (ACP) letter in response to a May 22 Senate Finance Committee letter requesting recommendations and policies that will improve care for patients with chronic diseases. In a 19-page letter ACP President Wayne J. Riley, MD, MBA, MPH, MACP, to Finance Committee Chairman Orrin Hatch; Finance Committee Ranking Member Ron Wyden; Co-chairman of the Chronic Care Working Group Johnny Isakson; and Co-chairman of the Chronic Care Working Group Mark Warner, provides 18 detailed recommendations to: Expand the flexibility of Medicare Advantage ...

Latina women undergoing chemotherapy for breast cancer need more stress management tools

2015-06-19
TAMPA, Fla. - Cancer is the most common cause of death among Latinos. It is estimated that one out of every three Latina women will be diagnosed with cancer during her lifetime. Given the increasing Latino population in the United States, more emphasis has to be placed on educating this population about cancer. Moffitt Cancer Center researchers, along with collaborators at the University of South Florida, recently published a study about the attitudes and cultural perspectives of Latinas undergoing chemotherapy for breast cancer. The article also discusses their cancer ...

NASA sees Tropical Depression Bill tracking through US

NASA sees Tropical Depression Bill tracking through US
2015-06-19
Tropical Depression Bill continues to be a soaker as it travels in an east-northeasterly direction from Arkansas toward the Ohio Valley. NASA's Aqua satellite and NOAA's GOES-East satellites provided a look at the extent and the movement of the storm. The MODIS instrument aboard NASA's Aqua satellite captured a visible image of Tropical Depression Bill's remnants over the central United States on June 18 at 19:45 UTC (3:45 p.m. EDT). At the time Aqua passed overhead, Bill's center was located in northeastern Oklahoma. The image showed that moisture from the Gulf of Mexico ...

LAST 30 PRESS RELEASES:

The move by Apple Memories to block potentially upsetting content illustrates Big Tech’s reach and limits, writes Chrys Vilvang

Chemical tool illuminates pathways used by dopamine, opioids and other neuronal signals

Asian monsoon lofts ozone-depleting substances to stratosphere

PET scans reveal ‘smoldering’ inflammation in patients with multiple sclerosis

Genetics predict type 2 diabetes risk and disparities in childhood cancer survivors

Health information on TikTok: The good, the bad and the ugly

New study points to racial and social barriers that block treatment for multiple myeloma

Rensselaer researcher finds that frog species evolved rapidly in response to road salts

A new chapter in quantum vortices: Customizing electron vortex beams

Don’t be a stranger – study finds rekindling old friendships as scary as making new ones

There’s no ‘one size fits all’ when it comes to addressing men’s health issues globally

Comparison of the “late catch-up” phenomenon between BuMA Supreme and XIENCE stents through serial optical coherence tomography at 1–2 month and 2 year follow-ups: A multicenter study

Marine plankton communities changed long before extinctions

Research reveals tools to make STEM degrees more affordable

Q&A: UW research shows neural connection between learning a second language and learning to code

Keane wins 2024 Gopal K. Shenoy Excellence in Beamline Science Award

Livestock abortion surveillance could protect livelihoods and detect emerging global pathogens

Optimal timing maximises Paxlovid benefits for treating COVID-19

IU researchers receive $4.8 million grant to study the role of misfolded protein TDP-43 in neurodegenerative diseases

DOE’s Office of Science Graduate Student Research Program selects 86 outstanding US graduate students

This tiny chip can safeguard user data while enabling efficient computing on a smartphone

World’s chocolate supply threatened by devastating virus

Wake up and die: Human brain neurons re-entering the cell cycle age quickly shift to senescence

Phage therapy is being explored to treat multidrug-resistant bacterial infections, but what are the direct effects of phages on the human host?

Social media use linked to tobacco initiation among youth

Marginalized communities developed 'disaster subculture' when living through extreme climate events, study finds

AGS honors Dr. William Hall with prestigious Nascher/Manning Award in Geriatrics

Human Frontier Science Program: life science research addressing sustainability of living systems

Wind turbine blades get a sustainable upgrade

New study uncovers lasting financial hardship associated with cancer diagnosis for working-age adults in the U.S.

[Press-News.org] Cardiac device wearers should keep distance from smartphones
Patients should hold phone to opposite ear and not store phone in pocket over device